panoramic dental
Technical data |
Compact control panel with intuitive symbols. |
• X-ray controls via keypad with spiral cable. |
•Optional remote control keypad. |
•Constant potential high-frequency multi-pulse X-ray generator. |
Focal spot: 0.5 IEC 336. |
Voltage at anode: 61 - 85 kV. |
Current at anode: 4 - 10 mA. |
Exposure time: panoramic 15 s maximum, cephalometric 9 to 12 s. |
Heating load management for minimum waiting time. |
Source image distance: panoramic 51 cm, cephalometric 165 cm. |
Dental arch vertical magnifi cation: 27% |
Cephalometric magnifi cation 10% |
Vertical travel of reference point for occlusal plane: 90 - 176 cm. |
Panoramic projections: (1) Adult standard with constant vertical magnifi cation |
• Cephalometric projections: (1) Antero Posterior (AP), (2) Latero Lateral (LL), |
• Accessories for positioning: bite block, chin rest, optional temporo-mandibular |
• Positioning references: 3 laser beams for the Frankfurt, mid-saggital |
• Carriage positioning: motorised horizontal travel, electromagnetic |
Weight: Pan Ceph 205 kg, Pan Solo 160 kg. |
• Room size: Minimum height 220 cm, maximum height 227 cm, |
|
Digitale receiver |
• Removable CCD Sensor |
Image height 22 cm |
Spatial resolution > 5 line pairs per mm |
Acquisition dynamics 12 levels |
• Imaging area of 15 x 30 cm on Panoramic and 22 x 24 cm on Ceph |
• USB interface |
digital panoramic dental x ray machine
Image