Detecting type | Electrochemical | Detecting Gas | Toxic and Hazardous gases | ||
Measuring Range | 0~25%VOL | Voltage Range | DC24V±10V | ||
Sampling | Diffusion | Power Consumption | ≤1.5W (DC24V) | ||
Communication | 4ABUS/ 3Wire 4~20mA | Communication range | ≤1500 m(1.5mm2) | ||
Ex CERTIFICATION | Exd Ⅱ CT6 | IP Rating | IP66 | ||
Enclosure material | Aluminum alloy. 304 Stainless steel .Tempered glass | Cable entries | G1/2 Internal screw thread | ||
Temperature-Humidity Range | -40℃~+70℃,≤93%RH | Operating Pressure Range | 86kPa~106kPa | ||
Operating | IR remote control >8m | Expected Operating Life | 3 years(Typcial) | ||
Alarm accuracy | ±3%VOL | Display accuracy | ±5%VOL | ||
Low alarm | 19.5%VOL(Factory setting) | High alarm | 23.5%VOL(Factory setting) | ||
Response time | <30s(T90) | Output range | ≤30V,2A(maximum limiting current) | ||
Use | 4~20mA 3Wire Communication | ||||
Output | 1 Passive switch output, associated with First/Second Limit . Remain opening in normal state. Alarming will get it closed. | ||||
Display type | 2.1 inch LED.4 Bit Nixietube. 8-Segment Numeric +Graphical display | ||||
Dimension | H196mm×L140mm×D92mm |