SYA-265B kinematic viscosity
Kinematic viscosity tester
Measuring device
265D ≥ 1, the viscosity range of the standard 7-branch capillary is 1.8 ≤ 2600, and the inner diameter of the standard 7 Pin capillaries is 0.60,0.81,1.01.2,1.50, 2.00.20. 2.5mm.The kinematic viscosity tester is designed and manufactured according to the GB/T265 standard, and can accurately carry out oil in strict accordance with the standard description.
Determination of the kinematic viscosity of the product. Suitable for determining the kinematic viscosity of liquid petroleum products (Newtonian liquids) in units of
Mm2/s. The exterior design is beautiful and generous, the structure design is reasonable, the operation is convenient, and the result is accurate.
SYA-265B kinematic viscosity
Kinematic viscosity tester
Measuring device
Performance description:
The kinematic viscosity measuring device uses a digital temperature control table to control the temperature of the constant temperature bath. The temperature control range is room temperature -100 ° C, and the temperature control accuracy is ± 0.1 ° C. The set temperature and actual temperature can be visually displayed, and temperature setting and temperature difference correction are easy. Equipped with 2 columns of special viscometer clips, the viscometer clip is designed ingeniously. Just insert the capillary viscometer into the viscometer clip, adjust the vertical position, and increase the accuracy of the measurement results to improve work efficiency. The constant temperature bath is a fully transparent glass cylinder, which is convenient for observing the flow of the sample and fully conforms to the standard description, which improves the accuracy of the measurement results.
SYA-265B kinematic viscosity
Kinematic viscosity tester
Measuring device
Technical Parameters:
1, input power: 220V ± 10% 50HZ
2, heater power: 1.8KW
3. Stirring motor: 15W1400/min
4, lighting: 9W energy-saving fluorescent lamp (pull series dedicated)
5, temperature control mode: single-chip digital display intelligent control
6, temperature control range: room temperature ~ 100 ° C
7, temperature control accuracy: ± 0.1 ° C
8, test hole position: 4 holes
9, ambient temperature: 5 ~ 35 ° C
10, relative humidity: <85% RH