● Safety Standard | IEC 61010-1 CAT.IV 600V IEC 61326 EMC specification: test, control and electrical equipment for testing IEC60529 IP64 (Outer closed state) |
● AC Power | 220V±10%,50/60 HZ ,20 VA |
● battery | 16.8V Lithium ion rechargeable battery |
● Battery Life | 10000V@100M, 4 hour |
● size | 30cm*30cm*35cm |
● weight | 6kg |
●Test voltage accuracy | 100% to 110% of the nominal value |
● Current test range | 10mA |
●Current measurement accuracy | 5%+0.2nA |
●Short-circuit current | 5mA |
●Capacitance test range | 20uF |
●Capacitance testing accuracy | 15%+0.03uF |
Insulation resistance testing range(T:23±5℃,rh;45-75%) | ||||||||
accuracy | 500V | 1000V | 2500V | 5000V | ||||
unspecified | <1M | <1M | <1M | <1M | ||||
5% | 1M-10G | 1M-20G | 1M-100G | 1M-100G | ||||
20% | 10G -100G | 20G-200G | 100G-1T | 100G-1T | ||||
unspecified | > 100G | > 200G | > 1T | > 1T |
Insulation resistance testing range(T:23±5℃,rh;45-75%) | ||||||||
accuracy | 10000V | 15000V | ||||||
unspecified | <1M | <1M | ||||||
5% | 1M-200G | 1M-200G | ||||||
20% | 200G-2T | 200G-2T | ||||||
unspecified | >2T | >2T |