Veterinary Multiparameter Patient monitor with Etco2 Parameter
. 12.1'' Color TFT Display
. Suitable for adult and pediatric and neonatal patient . Basic Parameters( ECG, SPO2, NIBP, TEMP,PR,RESP) in a durable case for bedside monitoring and transport . Audible and visual alarms with adjustable alarm ranges . Trend data 72 hours, 1000 groups of NIBP data,playback waveforms and data of 72 hours
Technical Specification: | ||||||
ECG: | ||||||
. Standard Leads: RA.; LA ; RL; LL ; V | ||||||
. Performance gains: 0.25 ; 0.5; 1 ; 2; (4 gears adjustable) | ||||||
. Scanning speed: 12.5mm/s,25mm/s,50mm/s | ||||||
. Working mode: Diagnosis, monitoring and operation | ||||||
. Protection: shock resistance, resistance to Electric knife | ||||||
HR: | ||||||
. Measure Range: 20bpm~250bpm | ||||||
. Alarm Range: Up:22bpm~250bpm;Lower:20bpm~248 bpm | ||||||
SPO2: | ||||||
.Measure range: 0%~ 100% | ||||||
.Alarm Range: Up: 70 % ~ 100% ; Low: 0% ~ 99% | ||||||
NIBP: | ||||||
. Measure Mode: Manu, Auto, STAT | ||||||
. Measuring range: systolic pressure: 50 mmHg ~ 240 mmHg mean blood pressure:25 mmHg ~ 200 mmHg diastolic pressure:15 mmHg ~ 180 mmHg | ||||||
.Alarm Range: systolic pressure:Up: 52 mmHg ~ 250 mmHg ; Low: 50 mmHg ~ 248 mmHg | ||||||
mean blood pressure:Up: 22 mmHg ~ 230 mmHg ; Low: 20 mmHg ~ 228 mmHg | ||||||
diastolic pressure:Up: 17 mmHg ~ 180 mmHg ; Low: 15 mmHg ~ 178 mmHg | ||||||
TEMP: | ||||||
.Measure Range: Not less than 28 ℃ ~43.2℃ .Alarm Range: Up: 0.2℃ ~ 50℃ ; Low: 0℃~ 49.8 ℃ | ||||||
PR: . Measure Range: Not less than 30bpm ~ 250bpm | ||||||
RESP: | ||||||
. Measure Range: Not less than 0 time/ min~ 100 times/ min | ||||||
. Display: Waveforms & Breathing Rate | ||||||
Optional: Patient monitor |