Specification
T:23°C±1°C;
H:40%~60%RH.;
Power supplier
Three phase:AC220/380V±15%;
Frequency:50Hz±5%
Class: 0.05
Output voltage:
Measuring range:3×57.7/100/220/380V
Setting range:0~120%
Setting accuracy:better than 0.01%
Output stability:≤0.05%/180s
Output degree of distortion:≤0.5%
Output capacity:>15VA/position
Load characteristic:resistance, inductive, capacitive
Capacitive load:4uF
Output current
Measuring range: 3×0.01A,0.1A,0.25A,0.5A,1A,2.5A,5A,10A,25A,50A,120A
Setting range:0~120%
Setting accuracy:better than 0.01%
Output stability:≤0.05%/180s
Output degree of distortion:≤0.5%
Output capacity:>25VA /possition
Load characteristic:Resistance, inductive
Start current output :min 1mA ,accuracy:≤5%
Output power:
Stability :≤0.05%/180s
Start power :accurancy≤ 5%;
Output phase position
Setting range :0~359.9°
Setting accuracy:0.01°
Output frequency:
Setting range:45~65Hz
Setting accuracy:better than 0.01 Hz
Output balance
a.Output voltage unbalance degree:≤0.2%
b.Output current unbalance degree:≤0.2%
Unit configuration:
Standard configuration for energy meter
Class:0.05
Multi-circuit voltage transformer
Class:0.01
Voltage: 60/120/240/480V
Current:1A/10A/120A
Meter position:3
Supervisory instrument
Voltage
Class:0.05%RD
Position:6
Resolution:0.001A
Current
Class:0.05%RD
Position:6
Resolution:0.0001A
Active power
Accuracy:0.05%RD
Reactive power
Accuracy:0.1%RD
Power factor
PF :-1.0~0~1.0
Accuracy:0.05%RG(PF:0.5-1)
position:5
Resolution:0.0001
Phase position(V/A)
Measuring range(θ):0-360°
Class:0.05°
Position:6
Resolution:0.001°
Frequency
Class:0.02%RD(45-65Hz)
Position:6
Resolution:0.0001Hz
Error processing system
1)equipped with internal error processing unit for each meter position
2)calculate and display error value, check and rest,
4)Protection for incorrect connection
RS485