Accuracy: +/-2% of FS at a constant
temperature; +/-5% of FS over
operating temperature range (once
temperature equilibrium has been
reached; except 0-1 ppm scale which
is +/-0.2 ppm)
Sensitivity: 0.5% of FS
Response time: 90% of FS at 77°F (25°C)
10 seconds for 0-1000 ppm or
higher ranges
60 seconds for 0-10 ppm range
100 seconds for 0-1 ppm range
Operating temp: 50° to 95°F (5° to 35°C)
Signal output: Analytical measurement 0-1VDC
and isolated 4-20mADC
Range ID output: 0-1VDC and isolated 4-20mADC
Max. load
impedance: 4-20mA isolated output -- 1000 ohms
Analysis display: 5 digit red LED, 3/5" high numerals
Menu display: 20 character, 2 line vacuum fluorescent
Data lines: Bi-directional RS-232C serial interface,
baud rate 2400
Power
requirements: Universal AC input ranges,
85 to 230VAC, 50 / 60 Hz
Max. power
consumption: 20 VA
Oxygen sensor: Micro-fuel cell, Class B-2CXL
Sample gas
inlet pressure: Min 2-3 psig
Wetted parts: 316 stainless steel
Sample inlet
connection: 1/4" VCR
Area
classification: General purpose
Mounting: Flush panel mount
Dimensions: 8.70" W x 6.96" H x 12.2" D (case)
10.79" W x 7.46" H (front panel)
MODEL 3000TA-XL High Sensitivity Trace Oxygen Analyzer
Regional Office
Tokyo, Japan ..............................................81-33-239-9090
SPARE PARTS - MODEL 3010TAC |
|
DESCRIPTION |
|
Back panel board |
|
Front panel board |
|
Trace preamplifier board |
|
Main computer board (standard) |
|
Fuse, 1A, 250V, 3AG, Slow Blow (US) |
|
Fuse, 1A, 250V, 5x20 mm, T-Slow Blow (EU) |
|
O-ring |
|
Micro-fuel Cell, class L-2C |
|
Intrinsic safety Zener barrier |
|
Backpanel connector kit |
|
Isolated 4-20mA module |
|
Cell spacer for B-2C or A-2C sensor |