parameters | Detail |
Infusion Principle | Fingertip Peristaltic Pump |
I.V.Set Compatibility | All brands, special or quality PVC, TPE infusion sets with an external diameter of 3.4mm~4.5mm |
Range of Flow Rate | 1~1200 ml/h |
Infusion Rate Step By | Increments at 0.1 ml/h |
Infusion Mode | Flow-rate mode, Time mode, Drop mode, Body weight mode |
Purge/Bolus Rate | Set between(50.0~800.0) ml/h for 0.1 ml/h step |
Time Preset | 1 s~99 h 59m 59 s) , step by 1 s |
Volume Range | 0.1~9999.99 ml |
Drop Range | 1-300 d/min, step by 1 drop |
Body Weight Mode | Drug amount range 0.1~999.9 mg, step by 0.1 mg |
Dosage range 0.01~999.99 mg/kg/h, step by 0.01 mg/kg/h | |
Weight range 1~300 kg, step by 0.1 Kg | |
Accuracy | ±5% |
Air-in-line Detection | Ultrasonic wave, and 5 grades for sensitivity adjustments |
Occlusion Pressure | For upstream occlusion, within-20 kPa~-60 kPa, and 8 grades for sensitivity adjustments |
For downstream occlusion, within 20 kPa~140 kPa, and 8 grades for sensitivity adjustments | |
Visible&Audible Alarms | Air inline, Upstream Occlusion, Down Stream Occlusion, Door open, VTB completion, Low battery, Drop error, Pause overtime, MPU error, Pressure detect module error, Motor malfunction, Low battery, Battery error, A.C Disconnection. |
Event Logo | 1000 event list including(Infusion start/stop time, flowrate, alarms etc) |
K.V.O rate | 1 ml/h~5 ml/h, defaulvalue1 ml/h, can be programmed by user step by 0.1 |
Internal Battery | Ni-Mh rechargeable battery, 12 V/800mAh, more than 3 hours working backup |
Power | AC:100 V-240 V 50 HZ/60 HZ |
Power Consumption | <20 VA |
Working Environment | Temperature:+5--+40°C |
Atmosphericpressure:20%--90% | |
RelativeHumidity:700 hPa--1060 hPa | |
Classification | Class I, Type CF, IPX 4 |
Dimension&Weight | 130 mm(L)x118 mm(W)x206 mm(H);1.4 Kg |
Optional Function | Drop Sensor, WIFI, Heating Function |