Technical Specifications
Measurement range |
0 % ... 100 % oxygen |
Nominal sensor lifetime |
≥ 1 000 000 % volume oxygen hours |
Output in ambient air |
13 mV ... 16 mV |
Electrical interface |
3 pin (Molex® 22-11-1031) |
Accuracy |
meets ISO 80601-2-55 requirements |
Response time |
< 12 s to 90 % of final value |
Influence of humidity |
-0.03 % rel. per % RH at 25 °C |
Operating temperature |
0 °C ... +50 °C |
Recommended load |
≥ 10 kOhms |
Weight |
approximately 28 grams |
Storage temperature |
-20 °C ... +50 °C |
Recommended storage |
+5 °C … +15 °C |
Operating humidity |
0 % ... 99 % RH non-condensing |
Repeatability |
< 1 % volume O2 at constant temperature and pressure < 3 % relative
|
Zero offset voltage |
< 200 µV in 100 % nitrogen, applied for 5 min
|
Cross interference |
meets ISO 80601-2-55 requirements |
Pressure range |
0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2) |
Warm-up time |
< 30 minutes, after replacement of sensor |