AC voltage (auto-ranging): 90-264 V rms, 47- 63 Hz 200 VA
Battery charge time: 2.5 hours deep discharge, 2 hours normal discharge
Battery life:11.1 V, 5.2Ah Li-ion batteries, meet IEC 62133:2003,
4.5 hours (typical) continuous testing at 15 kV with a 100 MΩ load
Test voltage:1000 V, 2500 V, 5000 V, 10000 V, 15000 V, VL
Test voltage accuracy:+4%, -0%, ±10 V nominal test voltage at 1 GΩ load (0°C to 30°C)
Resistance range:10 kΩ to 30 TΩ
Accuracy:±5% from 1MΩ to 3 TΩ; ±20% from 1MΩ to 30 TΩ
Guard terminal performance:Guards out parallel leakage resistance down to 250 kΩ with a maximum additional resistance error of 1% with a 100 MΩ load
Display
Analogue: 100 kΩ to 10 TΩ
Digital: 10 kΩ to max a above
Short circuit/charge current:3 mA
Capacitor charge:<7.5 s/μF at 3 mA to 15 kV
Capacitor discharge:<3500 ms/μF to discharge from 15000 V to 50 V
Capacitance range:0.01 nA to 6 mA
Capacitance measurement accuracy:±10% ±5 nF
Current range:0.01 nA to 6 mA
Current accuracy:±5% ±0.2 nA at all voltages (23 °C)
Interference:6 mA from 2100 V to 15 kV
Voltmeter range:30 V to 660 V ac or dc, 45Hz – 65Hz
Voltmeter accuracy:±3%, ±3V
Timer range:Up to 99 minutes 59 seconds, 15 second minimum setting
Memory capacity:5.5 hours logging @ 5 sec intervals
Maximum altitude:3000 m
Dimensions:L 360 mm x W 305 mm x H 194 mm
Weight:6.5kg