Project index | Parameter index | ||
Input Power | AC 220V ±10% / 50Hz | ||
resolution | AD:16bit;DA:16bit | ||
Input Resistance | ≥1MΩ | ||
Input Power | 3241W | ||
Channel Features | Constant Current Source and Constant Voltage Source with Independent pairs of Closed-loop Structure | ||
Channel Control Mode | Independent Control | ||
Voltage | Per Channel Voltage Range | Charge:0.15V~30V;Discharge:3.0V~30V | |
Accuracy | ± 0.1% of range(Full-scale) | ||
Stability | 0.1% | ||
Current | Per Channel Current Range | Charge:50mA~10A;Discharge:50mA~10A | |
Accuracy | ± 0.1% of range(Full-scale) | ||
Stability | 0.1% | ||
Power | Per Channel Output Power | 300W | |
Stability | 0.2% | ||
Time | Current Response Time | The Current gets to 10A Hardware Response Time is 20ms | |
Step Time Range | 1s≤Single Step Time Range≤365*24 h, Time Form Support 00:00:00(h,min,s) | ||
Data Recording | Record Condition | MinimumTime Interval:(1s ~60000s) | |
Minimum Voltage Interval:(0.15V~30V) | |||
Minimum Current Interval:(50mA~10A) | |||
Record Frequency | 10Hz | ||
Charge | Charge Mode | Constant Current Charge,Constant Voltage Charge, CC-CV Charge ,CP-Charge | |
End Conditon | Voltage,Current,Relative Time,Capacity | ||
Discharge | Discharge Mode | Constant Current Discharge CP-D CR-D | |
End Condition | Current,Current,Relative Time,Capacity | ||
Pulse Model | Charge | Constant Current,Constant Power | |
Discharge | Constant Current,Constant Power | ||
Min Pulse Width | 500ms | ||
Pulse Number | Single Pulse Step Support 32 different Pulse | ||
Charge r To Discharge | One Pulse Step Can Support Continuous Charge And Discharge Change | ||
Stop Condision | Voltage, Relative Time | ||
DCR Testing | Support Custom Access Point And Count The DCR | ||
Cycle | Loop Measure Range | 1~65535 times | |
Max Steps Per Loop | 254 | ||
Nested Loop | Nested loop Function,Max Support 3 Layers | ||
Protection | Software Protection | l Power-down Data Protection | |
l Setting Protection Condition,Setting Parameter:Low Voltage limit,Upper Voltage limit,Low Current limit,Upper Current limit,Delay Time | |||
Hardware Protection | Additional Anti-reverse Protection Model | ||
Voltage and Current Testing Sample | 4-wire connecting | ||
Noise | <80dB | ||
Database | MySQL Database | ||
A Means of Communication of Upper Machine | TCP/IP Protocol | ||
Data Output | EXCEL,TXT,Graph | ||
Communication Interface | Ethernet Port | ||
Per Unit Main Channels No. | 8 | ||
Dimensions | 18U,W*D*H:600*600*1000(mm) |
Project index | Parameter index |
Working Temperature Range | 0℃~40℃ |
Storage Temperature Range | -10℃~45℃ |
Working Environment Relative Humidity Range | 30% ~ 80% RH |
Storage Environment Relative Humidity Range | 30% ~ 90% RH |