Model | 900B (Advanced Model) | ||
Display | 10.4 inch high visibility color TFT display | ||
Ventilation Mode | Ventilation Mode | ||
Minute Volume | Greater than or equal to 18L | ||
Back-up Power Supply | At least 4 hours | ||
Tidal Volume | Adjustable range: 20-1500ml, Display range: 0-2000ml | ||
Respiratory Rate | 1-99bpm | ||
Inspiratory/expiratory (I:E) ratio | 4:1-1:4 | ||
PEEP | 0-10cmH2O | ||
SIMV Rate | 1-20bpm | ||
Inspiratory Pressure Trigger | .-10-10cmH2O | ||
Trigger Sensitivity (flow) | 1-10L/m | ||
Inspiratory Plateau | Adjustable from 0-50% of inspiratory time | ||
Adjustable Oxygen Concentration | 21-100% | ||
Gas Supply Requirement | 280~600 kPa medical grade oxygen and compressed air supply | ||
Pressure Limit | 1.0kPa~6.0kPa | ||
Maximum Safe Operating Pressure | ≤6.0kPa | ||
Tidal Volume Alarm | Upper Limit: 10~2000ml, Lower Limit: 0~1800ml | ||
Airway Pressure Alarm | Upper Limit: 0.1~6kPa, Lower Limit: 0~5kPa | ||
Oxygen Concentration Alarm | Upper Limit: 22~100%, Lower Limit: 21~80% |