Product display
Product parameters
1. Some Standard Configuration
Item |
Description
|
1 |
Mammography System |
|
a) Rating power:5000VA b) Integrated System (Generator Integrated within the main device) c) High voltage generator output: ≥100 kHz, 20 kV to 40 kV increment step: 1 kV d) High performance X-ray tube: Standard:Molybdenum Anode, Focal Spot Size: 0.1 mm (Small)/ 0.3 mm (Large) Optional:Tungsten Anode, Focal Spot Size: 0.1 mm (Small)/ 0.3 mm (Large) e) Support iso-center rotation function f) C-arm Rotation range: +180°~ -135° g) Height range of breast support surface: 750mm-1330mm; h) C-arm power-driven electrically, Dual side digital display; i) Compression plate: flexible and multi-level smart Compression j) Spatial Resolution: >6 Lp/mm k) Detector type: Amorphous Silicon Detector l) Effective Imaging Area: 24*30cm m) Acquisition Station compatible with DICOM 3.0 n) Focal distance : 650mm o) Added filter: Mo / Rh / AG p) Added filter Switch: Manual/Automatic q) X-ray Field adjustment: Manual r) Magnification factor: 1.5 s) Exposure mode: Manual / AEC t) Supply: 220V~,±10% u) Frequency: 50/60Hz±1Hz v) Max. power : 5000 VA w) N.W.: 240 Kg x) Compact design, minimum device height: < 1.1 m y) Packing Size: 182 X 103 X 139 CM |
2 |
High-Voltage Generator |
|
a) Frequency:≥100KHz; b) Tube voltage Range: 20 kV ~ 40 kV in 1 kV step; c) Max Output: 4 kW; d) Max tube voltage:40kV; e) Max tube current: 140mA; f) Max current time product: ≥630mAs; g) Ripple: <4%.
|
3 |
X-Ray Tube |
|
Standard: a) Imported X-ray tube for Mammography; b) Anode Type: Molybdenum; c) Focal Spot Size: 0.1 mm (Small)/ 0.3 mm (Large) d) Permanent filtration: 0.5mmBe e) Max anode speed: 10000rpm/min; f) Max tube voltage:40kV; g) Max tube current: 35mA(small)/140mA(Large) h) Maximun tube assembly heat content: 320 KJ |
Optional: a) Imported X-ray tube for Mammography; b) Anode Type: Tungsten; a) Focal Spot Size: 0.1 mm (Small)/ 0.3 mm (Large) b) Permanent filtration:0.5mmBe; c) Max anode speed: 10000rpm/min; d) Max tube voltage:49kV; e) Max tube current: 35mA(small)/140mA(Large) f) Maximun tube assembly heat content: 320 KJ |
|
4 |
Beam Limiting Device |
|
a) Electric switch between Molybdenum or Tungsten or Argentum added filter; b) Adjustable X-ray field manually. c) Light field indicator is on automatically as compression plate moves down; d) Light field indicator is off with delay as compression plate stops moving; e) Off delay of light field indicator can be set via software. |
5 |
C-arm Assembly |
|
a) C-arm up-down and rotation movement controlled via C-arm control panel; b) C-arm stops automatically at common mammographic position: CC, OBL and LAT; c) Rotation angle of OBL position can be set via software; d) Support table up-down range: 750mm(L)~1330mm(H); e) Rotation type of C-arm: isocenter rotation around the examined part; f) Rotation range of C-arm: -135o~180o g) Distance between focal spot and image reception area:650mm; h) Magnification factor:1.5 i) Device can be installed in the movable physical vehicle as max height of device is less than 110cm; j) Local display panel on the both side of the stand can indicate rotation angle of C-arm; k) For safer operation, emergency stop button is locate on the top of C-arm; l) Duplex foot switch(2 pcs) provided. |
6 |
Compression Device |
|
a) Up-down movement : power-driven electrically; b) Compression control: continuous activation by foot; c) Compression mode: flexible and multi-level compression, automatic and manual decompression; d) Compression plate travel range: 5~268mm; e) Compression force: 0~200N f) Compression thickness: 5~268mm g) Size of compression plate: 24X30cm. |