item | value |
Warranty | 1 Year |
Customized support | OEM |
Place of Origin | Sichuan China |
Brand Name | RuiCe |
Model Number | RCKS-3 |
Power | Electronic |
Temperature measurement range | -59.9℃-399.9℃ |
Temperature control range | Room temperature -399.9℃ |
Repeatability | ≤150℃ 3℃ >150℃ 5℃ |
Resolution | 0.1℃ |
Accuracy | 0.5% |
Temperature measurement sensors | Platinum resistance (PT100) |
Flash Fire Sensor | Ion detection ring |
Ambient temperature | 10-40℃ |
Dimension | 440X280X290(mm) |
Instrument weight | 14kg |
Phenomenon | Reason | Processing | |
1 | Turn on the power, LCD screen No display Phenomenon | ●The power supply is not plugged in properly. ●Each plug is loose. | Plug in the power Plug in the plug with loose |
2 | Large specimen repeatability error | ●Stirring shaft broken Thermocouple with oil affects sensitivity | Replace the stirring shaft Dry the oil on the thermocouple with filter paper |
3 | No heat | ●Electric heating wire break | Replacement of electric heating wire |
4 | Lifting arm does not rise or fall | ●Motor and lift arm screw off ●Damaged limit optocouple | Tighten the screws Replace the limit optocoupler |