General Technical Specifications
l Environment condition of using:
600V CAT.IV and 1000V CAT.III
Pollution level: 2
Altitude < 2000m
Working environment temperature and humidity: 0~40°C (<80% RH, <10°C non condensing).
Storage environment temperature and humidity: -10~60°C (<70% RH, remove the battery).
l Temperature coefficient: 0.1´ accuracy/°C (<18°C or >28°C).
l MAX. Voltage between terminals and earth ground:1000V DC or 750V AC RMS
l Fuse protection: mA: fuse F600mA/1000V fast speed fuse; 10A shift: F10A/1000V fast speed fuse
l Sampling rate: about 3 times/second.
l Display: 6000 counter readout. Automatically display the unit symbols according to the shift of the measurement function.
l Over range indication: it displays “OL”.
l Low battery indication: when the battery voltage is lower than the normal working voltage, “” will be displayed.
l Input polarity indication: automatically display “-”.
l Power requirement: 4 x 1.5V AA batteries.
l Dimension: 204(L) ´94(W) ´ 57(H) mm.
l Weight: about 410g (including batteries).
Accuracy Specifications
The accuracy applies within one year after the calibration.
Reference condition: the environment temperature 18°C to 28°C, the relative humidity is no more than 80%, accuracy: ±(% reading + word).
DC Voltage
Range | Resolution | Accuracy | Input impedance: 10MΩ(60/600mV range is 1011Ω) Overload protection: 1000V DC or 750V AC (RMS) Maximum input voltage: 1000V DC Note: Measure in mV (60/600mV) voltage, input impedance is as high as 1011Ω, and weak signal of measurement does not weaken, so the measurement precision is high. But in the circumstance that the probe is open circuit it will display numbers, which is normal, as long as in short circuit the probe will return to zero or the probe is inserted to the measured point, the reading will stabilize. |
60mV | 0.01mV | ±(0.7% reading+2) | |
600mV | 0.1mV | ||
6V | 1mV | ||
60V | 10mV | ||
600V | 100mV | ||
1000V | 1V |
Range | Resolution | Accuracy | Input impedance: 10MΩ(60/600mV range is 1011Ω) Overload protection: 1000V DC or 750V AC (RMS) Maximum input voltage: 750V AC (RMS) Frequency range: 10Hz ~ 1kHz; response: true RMS Note: Measure in mV (60/600mV) voltage, input impedance is as high as 1011Ω, and weak signal of measurement does not weaken, so the measurement precision is high. But in the circumstance that the probe is open circuit it will display numbers, which is normal, as long as in short circuit the probe will return to zero or the probe is inserted to the measured point, the reading will stabilize. |
60mV | 0.01mV | ±(0.8% reading+3) | |
600mV | 0.1mV | ||
6V | 1mV | ||
60V | 10mV | ||
600V | 100mV | ||
750V | 1V | ±(1.0% reading +3) |
Range | Resolution | Accuracy | Overload protection: mA/mA range:600mA/1000V fuse (fast speed fuse) 10A range: 10A/1000V fuse (fast speed fuse) Maximum input current: mA socket: 600mA 10A socket: 10A When measuring large current, continuous measurement should be not longer than 15 seconds, after measurement the instrument should be cooled down after twice of the measurement time, and then enter small current measurement. |
600mA | 0.1mA | ±(1.2% reading +3) | |
6000mA | 1mA | ||
60mA | 10mA | ||
600mA | 100mA | ||
6A | 1mA | ±(2.0% reading +10) | |
10A | 10mA |
AC Current
Range | Resolution | Accuracy | Overload protection: mA/mA range: 600mA/1000V fuse (fast speed fuse) 10A range: 10A/1000V fuse (fast speed fuse) Maximum input current: mA socket: 600mA 10A socket: 10A Frequency range: 10Hz~1kHz Response: true RMS When measuring large current, continuous measurement should be no longer than 15 seconds, after measurement the instrument should be cooled down after twice of the measurement time, then enter small current measurement. |
600mA | 0.1mA | ±(1.5% reading +3) | |
6000mA | 1mA | ||
60mA | 10mA | ||
600mA | 100mA | ||
6A | 1mA | ±(3.0% reading +10) | |
10A | 10mA |
Range | Resolution | Accuracy | Open circuit voltage: about 0.25V Overload protection: 1000V DC or 750V AC (RMS) |
600Ω | 0.1Ω | ±(1.2% reading+5) | |
6kΩ | 1Ω | ||
60kΩ | 10Ω | ||
600kΩ | 100Ω | ||
6MΩ | 1kΩ | ||
60MΩ | 10kΩ | ±(2.0% reading+5) |