Specifications
* Model: FSQ-M-VMX1
* Power Supply: 5kW
* Output Voltage: 20 - 40kV
* Input Voltage: 220V
* Range of Current: 10-200mA
* Range of mAs: 10 - 120mAs (Small focus)
* Range of mAs: 10 - 500mAs (Large focus)
* Frequency: 50/60KHZ
* Work Mode: Intermittent
* Power Requirement: 180-264V AC
X-Ray Tube
Specifications
* Model: QG-M2
* Focus: 0.1 / 0.3mm
* Anode Heat Storage Capacity: 300kHU
* Target Surface: RT-TZM Target
* Anode type: Rotating
X-ray Imaging System
Specifications
* Model: CareView 750M
* Detector Technology: a-Si Flat Panel Detector
* Scintillator Material: CsI
* Detector Size: 240 x 300mm
* Static Pixel Matrix: 4096 x 3072pixels
* Pixel Size: 77μm
* Grayscale: 16 bit
* Preview Image: ≤ 5s
* Spatial Resolution: 6 lp/mm
* AEC Function
C-Arm Stand
Specifications
* Model: DM156-1
* FID: 1420 - 2020mm
* Angle Range of Arm Rotation: -135° - +180°
* SID: 650mm
* Range of Arm Up and Down: 0—600mm
* One-click Function
* Isocentric motion
Compression Device
Specifications
* Model: YPQM-1/YPQM-2/YPQM-3
* Maximum Height: 280mm
* Compressor Size: 240mm × 310mm (YPQM-1)
* Compressor Size: 100mm × 100mm (YPQM-2)
* Compressor Size: 200mm × 280mm (YPQM-3) (Optional)
* Pressure: 0 – 196N
* Compressor: Automatic Release Function and Manual Control Function
Grid
Specifications
* Model: JPI 2100
* Grid Density: 332.5 Line/Inch(±1)
* Grid Ratio: 4 : 1
* Focus: 645mm
Collimator
Specifications
* Model: XSM-2
* Filter: 0.1mm Al (28KV)
DX Ray Diagnostic Workstation
Specifications
* Model: TSM-1
* CPU: Dual-Core 3.7G
* RAM: 4G
* Hard Disk: 500G
* Display: 2M Medical Grade & 24-inch LCD Monitor
* Registration management, image acquisition, image viewer, image processing, film printing, system settings
* DICOM connection, original data, DICOM data storage
* Distance measurement, angle measurement, adjustment of image contrast and brightness , image rotation, image flip, negative image, image distortion correction
* Supports Standard DICOM Printer