Model | BK-6190 | ||||||||
Throughput | 60Tests/hour | ||||||||
Measuring Principle | Impedance for cell counting Cyanide-free method for HGB | ||||||||
Aperture Diameter | WBC chamber:100μm,RBC/PLT chamber:70μm | ||||||||
Parameters | WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR | ||||||||
Histograms | 3 histograms(includingWBC, RBC, PLT histogram) | ||||||||
Sample Volume | Normal:10μl; Predileted:20μl | ||||||||
Calibration | Manual & auto calibration | ||||||||
Counting Modes | Venous, capilary, prediluted | ||||||||
Printer | Built-in thermal printer,USB to external printer(optional) | ||||||||
Data Input | 8 inch touch screen, keyboard, mouse | ||||||||
Output | RS232 port, USB port, VGA port. LIS system avaliable | ||||||||
Storage | 100,000 results with histogram | ||||||||
Carry-over rate | Parameter | CV | |||||||
WBC | ≤ 2% | ||||||||
RBC | ≤ 1% | ||||||||
HGB | ≤ 2% | ||||||||
PLT | ≤ 2% | ||||||||
Background | Parameter | Background | |||||||
WBC | ≤ 0.2×10^9/L | ||||||||
RBC | ≤ 0.02×10^12/L | ||||||||
HGB | ≤ 1 g/L | ||||||||
HCT | ≤ 0.5% | ||||||||
PLT | ≤ 10×10^9/L | ||||||||
Parameter | CV | Measurement range | |||||||
Precision | WBC | ≤ 2.0% | (4.0-15.0×10^9)/L | ||||||
RBC | ≤ 1.9% | (3.5-6.0×10^12)/L | |||||||
HGB | ≤ 1.9% | (110.0-180.0g)/L | |||||||
MCV | ≤ 0.4% | (80.0-110.0)fL | |||||||
PLT | ≤ 4.0% | (100.0-500.0×10^9)/L | |||||||
Linearity | Parameter | Measurement range | CV | ||||||
WBC | (0-5.9×10^9)/L | (±0.3×10^9)/L | |||||||
(6.0-99.9×10^9)/L | ±5% | ||||||||
RBC | (0-0.99×10^12)/L | (±0.05×10^12)/L | |||||||
(1.0-9.99×10^12)/L | ±5% | ||||||||
HGB | (0-99.0g)/L | (±2.0g)/L | |||||||
(100.0-300.0g)/L | ±2% | ||||||||
PLT | (0-99×10^9)/L | (±8.0×10^9)/L | |||||||
100-999×10^9)/L | ±10% | ||||||||
Power supply | AC 220±10%,60/50Hz, 110±10%,60Hz |